Title: | Compatibility determination of potassium orotate with spironolactone by high-performance liquid chromatography |
Authors: | Mazur, Ecaterina Schmid, Martin Uncu, Livia |
Keywords: | HPLC;combination;potassium orotate;spironolactone |
Issue Date: | 2020 |
Publisher: | The Scientific Medical Association of the Republic of Moldova |
Citation: | MAZUR, Ecaterina, SCHMID, Martin, UNCU, Livia. Compatibility determination of potassium orotate with spironolactone by high-performance liquid chromatography. In: The Moldovan Medical Journal. 2020, vol. 63, no 4, pp. 35-42. ISSN 2537-6381. DOI: 10.5281/zenodo.4016810 |
Abstract: | Background: Compatibility determination between active pharmaceutical ingredients (APIs) in fixed-dose combinations is an indispensable step in the
elaboration. High-performance liquid chromatography provides information on possible interactions between APIs and their related interaction products.
The purpose of the present study was to investigate the compatibility of potassium orotate in combination with spironolactone by a HPLC method.
Material and methods: The detection was carried out using Liquid Chromatograph Agilent 1100 with UV-VIS detector and a RP-18 reversed column
(250*4 mm, 5 μm), mobile phase of acetonitrile: phosphate buffer solution (pH=4.0) with the ratio 1:49 and 1:1, at flow rate 1 and 1.5 mL/min, injection
volume 20 μL; potassium orotate and spironolactone substances were provided by Sigma Aldrich, USA.
Results: Due to the developed method both separation and simultaneous qualitative and quantitative determination of APIs in the mechanical mixture
were carried out. Spironolactone: retention time 6.9 min, concentration 98.1% (±0.21); potassium orotate: retention time 3.06 min, concentration 91.67%
(±0.15). There were just well-separated symmetrical peaks of APIs and no additional peak in the chromatograms.
Conclusions: There is compatibility between APIs. Further studies will be performed by other methods (DSC, FT-IF Spectrometry) to confirm the
obtained result. |
metadata.dc.relation.ispartof: | The Moldovan Medical Journal: The 75th anniversary of Nicolae Testemitanu State University of Medicine and Pharmacy of the Republic of Moldova (1945-2020) |
URI: | https://doi.org/10.5281/zenodo.4016810 http://moldmedjournal.md/wp-content/uploads/2020/09/MMJ-Vol-63-No-4-Oct-2020.pdf http://repository.usmf.md/handle/20.500.12710/11730 |
ISSN: | 2537-6381 2537-6373 |
Appears in Collections: | The Moldovan Medical Journal, Vol. 63, No 4, October 2020
|
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.
|