Title: Development of a UV-VIS spectrophotometric Azithromycin assay
Authors: Anton, Mihail
Vislouh, Oxana
Issue Date: 2010
Publisher: Nicolae Testemitanu State Medical and Pharmaceutical University
Citation: ANTON, Mihail, VISLOUH, Oxana. Development of a UV-VIS spectrophotometric Azithromycin assay. In: MedEspera: the 3rd Internat. Medical Congress for Students and Young Doctors: abstract book. Chișinău: S. n., 2010, p. 103.
Abstract: The goal of project was the elaboration of an alternative Azithromycin assay given the fact that the Pharmacopoeial method is the biological one. This method is hardly realizable, and requires special determination conditions, which also creates some difficulties in the substance and pharmaceutical forms analysis. Materials and methods: For spectral analysis Agilent - 8453 UV-VIS spectrophotometers in wavelengths range from 350 up to 550 nm were used. Phosphate buffer solutions with different pH values were used as solvent. The colored product, resulting from reaction with concentrated sulphuric acid, was investigated. Results: Azithromycin, which is a compound with complex structure and is part of macrolide antibiotics group, forms a colored compound in reaction with concentrated sulphuric acid. In order to determine the possibility of using this compound in the UV-VIS spectrophotometric assay, we have studied the properties of mixture complex being composed from Azithromycin and Sulphuric acid; such as parameters also were assayed: the stability over the derivatization time, its dependency on reagents co-ratio and solvent’s influence, and the value of pH environment. There were been developed the optimal technique of life time colouring reaction: the amount of reagents - 10 ml; reaction time - 30 min; pH - 7.0. The spectrum of absorption was recorded in the range from 350 up to 550 nm; the maximum of absorption was established at 483 nm. The basic parameters of the method were determined: solution concentration, specific absorption, absorbance law enforcement. The linearity and accuracy of the method were evaluated that range within the limits of about 3%. Conclusions: 1. There were been established the optimum conditions of Azithromycin assay, using the UV-VIS spectrophotomety on the base of coloring reaction of Azithromycin with concentrated Sulphuric acid. 2. There were been developed valuable technique and have evaluated the main criteria of validation: linearity, accuracy and the repeatability of the method. 3. UV-VIS spectrophotometry of quantitative Azithromycin determination may be also employed in the determination of this active principle in the pharmaceutical forms.
metadata.dc.relation.ispartof: MedEspera: The 3rd International Medical Congress for Students and Young Doctors, May 19-21, 2010, Chisinau, Republic of Moldova
URI: http://repository.usmf.md/handle/20.500.12710/20327
Appears in Collections:MedEspera 2010

Files in This Item:
File Description SizeFormat 
Development_of_a_UV_VIS_spectrophotometric_Azithromycin_assay.pdf635.62 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

 