Title: | Spectrophotometric dosing of beta carotene in tablets method validation |
Authors: | Drumea, Maria Podgornîi, Ana |
Keywords: | validation;Spetrophotometry UV-VIS;Eurosept;b-Carotene |
Issue Date: | 2014 |
Publisher: | Ministry of Health of the Republic of Moldova, State Medical and Pharmaceutical University Nicolae Testemitanu, Medical Students and Residents Association |
Citation: | DRUMEA, Maria, PODGORNÎI, Ana. Spectrophotometric dosing of beta carotene in tablets method validation. In: MedEspera: the 5th Internat. Medical Congress for Students and Young Doctors: abstract book. Chișinău: S. n., 2014, p. 238. |
Abstract: | Introduction: Validation is a verification methodology, compliance and accreditation of
validity as a method of scientific analysis. This methodology aims to demonstrate that the analysis
method corresponds to the utilization for which it was for seen, and to the parameters of validation
in accordance with regulatory organizations. It is now recognized that no drug substance or drug
can enter the European and American market without being accompanied by corresponding
validation folder. The results of this study will serve as support in concluding validation methods
report for the determination of the active principles of EUROSEPT product.
Materials and methods: The study was conducted on lozenges EUROSEPT, series 041211
made in December 4th, 2012 and valid until 12.2014. 1 tablet contains: active substances: ascorbic
acid, eucalyptus essential oil, Tolu balsam, P-carotene; auxeliary substances: aerosil, corn starch,
talc, magnesium stearate, stearic acid, citric acid, aspartame, sorbitol, lactose. In the process of
dosage methods validation the following parameters were analyzed: specificity, precision
(repeatability, reproducibility), linearity, accuracy. Spectrophotometric dosage method was
performed by using scanning spectrophotometer double perchin Elmer Lambda 25 UV-VIS.
Results: Comparative analysis performance of the validation parameters were done,
according with the European and international regulatory organizations: ISO, FDA, ICH, USP.
Empirical results expressed through absorbance value were used for reconstruction of five
calibration graphs used for the completion of the final one. The correlation coefficient of linear
regression model is 0.9997, which shows that the experimental points approximate and arrange
around the regression line according to experimental conditions stipulated in the MFT. The
calibration graph is an average of all data points, which is characterized by a relative error less than
2%. Standard deviation of residuals, standard deviation of intercept and slope values are less than
0.5%, which indicates that the calibration chart can be used to determine the specificity of the
dosage method. At the maximum absorption of 454 nm, for the solution obtained in the placebo controlled trial were not observed significant absorbance values, which indicates that the secondary
constituents do not interfere the procedure of the active substances quantification (ascorbic acid,
eucalyptus oil, balm Tolu) of the reconstitude pharmaceutical form. Statistical obliquity value does
not exceed the 1.5%, which demonstrates the accuracy of the spectrophotometric dosage method of
beta-carotene in the visible domain. The correlation coefficient r2 is 0.9984, which indicates that
with the change in the amount of beta-carotene concentration of 80-120%, the dependence of
absorbance versus concentration signal is one almost linear.
Conclusion: Spectrophotometric UV-VIS validation method of b-carotene dosage within the
preparation EUROSEPT was performed, according to validation parameters: linearity, repeatability,
reproductibility, specificity, accuracy, according to regulatory organizations. |