| Title: | FAILURE RATES OF INTERBODY SPINAL CAGES BASED ON MATERIAL: A COMPREHENSIVE MAUDE DATABASE STUDY |
| Authors: | Ungurean Jr, Victor |
| Keywords: | interbody cages;polyetheretherketone;PEEK;MAUDE;cage failure |
| Issue Date: | 2026 |
| Publisher: | Centrul Editorial-Poligrafic Medicina |
| Citation: | Ungurean Jr, Victor. FAILURE RATES OF INTERBODY SPINAL CAGES BASED ON MATERIAL: A COMPREHENSIVE MAUDE DATABASE STUDY. In: MedEspera 2026, p. 412. ISBN 978-9975-82-486-6. |
| Abstract: | Introduction. Although interbody cages are one of the most popular solutions for spinal fusion, implant failure has been observed as a recurring event throughout all stages of spinal surgery. These events are underreported worldwide due to the absence of a standardized database for implant malfunctions. Various interbody cage materials such as polyetheretherketone (PEEK), titanium, silicon nitride, and carbon fiber-reinforced polymer have been reported to have specific advantages across the literature. However, few studies to date have directly compared these materials in large samples extrapolated from nationwide databases, with current literature being limited to case studies. Aim of study. This study aims to assess the failure rates of interbody cages by material type using a large -scale database, allowing spine surgeons to optimize implant selection for their patients. Methods and materials. The Food and Drug Administration’s (FDA) Manufacturer and User Facility Device Experience (MAUDE) database was queried to yield a comprehensive list of all interbody cage failures reported from 2012 to 2025. Subsequent analysis was performed to categorize the failure reports based on material and failure type. The reports were categorized by material (polyetheretherketone [PEEK], titanium, silicon nitride, and carbon fiber-reinforced polymer) and failure type (material deterioration, cage breakage, assembly integrity failure of both cage and instrumentation, and implant migration). Ultimately, the aggregated data was analyzed using descriptive statistics to determine the total prevalence of failure based on material type. Results. A total of 2,455 interbody cage failure reports were analyzed. Cage breakage was the most common failure type, accounting for approximately two-thirds of all reported malfunctions. PEEK represented the majority of reported failures (90%), followed by titanium (4%) and silicon nitride (<1%). Instrumentation failure represented 13% of all reports, followed by cage migration (9%) and assembly integrity failure (8%). Titanium cages were involved in 58% of all cage migration reports, followed by PEEK (42%) and silicon nitride (1%). Carbon fiber-reinforced polymer showed no failure events. Conclusions. PEEK interbody cages demonstrated the highest rates of breakage among all materials and comparable rates of migration to titanium cages. This is significant given the lower relative popularity of PEEK when compared to the gold standard, titanium. While titanium accounted for the highest net number and proportion of migrations, these rates are likely skewed by its widespread prevalence among the materials used for the manufacturing of interbody cages. Ultimately, the findings in this study highlight the importance of implant material selection before surgery to maximize patient outcomes and reduce revision rates. |
| metadata.dc.relation.ispartof: | MedEspera 2026. The 11th International Medical Congress for Students and Young Doctors, 6-8 May 2026 |
| URI: | https://medespera.md/documents/abstract-book-4.pdf https://repository.usmf.md/handle/20.500.12710/34228 |
| ISBN: | 978-9975-82-486-6 |
| Appears in Collections: | MedEspera 2026
|
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.
|