Title: | Development and validation of the high-pressure liquid chromatography method for propyltiohinotiadiazol dosage |
Authors: | Uncu, Andrei |
Keywords: | HPLC;propylthiohinothiadiazole;dosing;validation |
Issue Date: | Oct-2020 |
Publisher: | Universitatea de Stat de Medicină şi Farmacie "Nicolae Testemiţanu" din Republica Moldova |
Abstract: | INTRODUCTION. Propylthiohinothiadiazole is a thiadiazole
derivative with pronounced antimycobacterial
action. Substance is researched in order to develop
methods of analysis and standardization, analytical
documents regulating quality for use as an active
principle in the development of pharmaceutical
forms.
The purpose. Elaboration of the work technique
and validation of the HPLC method for dosing
propylthiohinothiadiazole. MATERIAL AND METHODS. 3 series of propylthiohinothiadiazole, internal
reference standard of 2-(propylthio)-5H-[1,3,4]-
thiadiazole[2,3-b]-quinazolin-5-one substance with
concentration 99,98%, chromatographic system Shimadzu
LC-20AD HPLC with UV-VIS detector, acetonitrile and
dimethylsulfoxide of purity grade “pro HPLC analysis”
(Sigma Aldrih and Merck), bidistilled purified water.
RESULTS. Linearity is proved for concentrations of 5-30
µg/ml, the linear regression equation is y=36472x44580, R²=0.9929. The limit of detection is 0,729
µg/ml and the limit of quantification is 2,210
µg/ml. It was established that the method is
accurate (mean recovery values at 80%, 100% and
120% concentration levels were close to 100%. The
accuracy of the method was expressed by
repeatability and intermediate accuracy. The
variation of the chromatographic conditions
established that the method is robust. For all
validation parameters, RSD was less than 1. CONCLUSIONS. The validation results show that the developed HPLC method is simple, fast, accurate and reproducible. |