Abstract:
Introduction/Objectives. Optimization of transscleral cyclophotocoagulation
requires an objective assessment of tissue thermal response, as maximum
temperature alone does not adequately characterize treatment efficacy and safety.
This study aimed to develop an integrated system for quantitative evaluation of
tissue thermal response based on mathematical modeling.
Materials and Methods. A mathematical heat-transfer model of the ciliary body
during 810-nm transscleral cyclophotocoagulation was developed to simulate the
spatiotemporal distribution of tissue temperature under different laser settings.
Based on the simulated thermal profiles, an integrated system of thermal indices
was proposed, including the Thermal Efficiency Index (TEI), Destructive Energy
Index (DEI), Thermal Selectivity Index (TSI), Safety Margin, and Thermal
Accumulation Index (TAI).
Results. The proposed integrated indices provided a comprehensive quantitative
assessment of therapeutic efficiency, laser selectivity, thermal accumulation, and
the risk of destructive thermal injury. Unlike conventional evaluation based solely
on peak temperature, the integrated indices characterized the entire
spatiotemporal thermal profile during laser exposure. Laser settings of up to 2.0
W achieved the therapeutic temperature range, whereas increasing energy
delivery was associated with progressive thermal accumulation and a higher risk
of destructive tissue heating. The integrated assessment enabled identification of
the transition between therapeutic and potentially destructive thermal effects.
Conclusions. The proposed integrated thermal assessment system expands the
capabilities of mathematical modeling for transscleral cyclophotocoagulation by
providing an objective evaluation of treatment efficiency and safety. This
approach may support personalized optimization of laser parameters and improve
clinical planning of glaucoma treatment.